Line 28: Line 28:
 A critical component when using this methodology is determining the percentage of the maximum permissible error. The higher the percentage, the greater the risk of an instrument being found Out-of-Tolerance (OOT); potentially resulting in nonconforming product escapes. The lower the percentage, the greater the cost associated with lowering the risk of an OOT condition; and reducing the potential for nonconforming product escapes. This percentage will often vary based upon the type of instrumentation to which it is applied. \\ A critical component when using this methodology is determining the percentage of the maximum permissible error. The higher the percentage, the greater the risk of an instrument being found Out-of-Tolerance (OOT); potentially resulting in nonconforming product escapes. The lower the percentage, the greater the cost associated with lowering the risk of an OOT condition; and reducing the potential for nonconforming product escapes. This percentage will often vary based upon the type of instrumentation to which it is applied. \\
  
-Most often companies establish a "range" (or "window) for the optimization. For example, IF an instrument is found exceeding 75% of its maximum permissible error, then the calibration interval is shortened. However, IF an instrument is consistently found below 50% of its maximum permissible error, then the calibration interval is lengthened. And IF the instrument is found between 50% and 75% of it's maximum permissible error, then the interval is considered acceptable.+Most often companies establish a "range" (or "window") for the optimization. For example, IF an instrument is found exceeding 75% of its maximum permissible error, then the calibration interval is shortened. However, IF an instrument is consistently found below 50% of its maximum permissible error, then the calibration interval is lengthened. And IF the instrument is found between 50% and 75% of it's maximum permissible error, then the interval is considered acceptable.
 ===== Initial Calibration Intervals ===== ===== Initial Calibration Intervals =====